HbA1c Analyzer POCT HbA1c Analyzer LD-100 One Step Automatic Hematology Detection Analyzer
Intended Use
- It is intended for the percent determination of hemoglobin A1c (HbA1c) in human whole blood.
- HbA1c Analyzer is a semi-automatic instrument used to analyze glycated hemoglobin.
- It has the characteristics of small sample consumption and easy operation.
- It is suitable for the detection of HbA1c in medical units at all levels.HbA1c Analyzer is composed of the systems of sample cup mixing, colorimetric detection and system control, result display.
Specifications
Model | LD-100 |
Method | Boronate Affinity |
Operating Type | 1-Step, fully automatic |
Measurement Time | Less than 3 minutes |
Sample Type | Whole blood |
Measurement Range | HbA1c 4.0~15%(NGSP Unit) eAG 68-385mg/dl(mmol/l) |
Operating Condition | Temperature:15~35℃ Humidity:15~80%R.H |
Storage Condition | Analyzer:10-60℃; Test kit: 2~8℃ |
Shelf-Life of Test Kit | 9 months |
Voltage | 12VDC-3.33A |
Sample Volume | 4ul |
Power connection | Detachable power supply cord |
Device category | Fixed |
Operating mode | Continuous |

PRINCIPLE
- Uses the (borate affinity) fluorescent immunoturbidimetry
- Push a mixture containing a fluorescent borate conjugate into an integrated test cup and mix to dissolve sufficiently.
- The blood samples in the sample collector are then pushed into the buffer and mixed to release the hemoglobin from the red blood cells.
- The total concentration of hemoglobin depends on the initial reduced fluorescence signal.
- Fluorescent borate conjugates bind to glycosylated hemoglobin, and glycosylated hemoglobin is determined by detecting the decrease of fluorescent active ingredients.
- The percentage of glycosylated hemoglobin in total hemoglobin can be determined.
Feature
- One-step Operation
- Boronate Affinity
- Battery Integrated
- Portable, Compact
- Cv<3%, Excellent Correlation With HPLC
- RFID For Batch Management
- Color touch screen
- External printer (Blue-tooth)
- USB Port
- LIS Support
Reliable
IFCC calibrator | Normal | Abnormal | Normal | Abnormal |
Mean HbA1c% | 5.4 | 9.6 | 5.3 | 9.2 |
CV (%) | 2.5% | 1.9% | 2.4% | 1.8% |
Use Step
- Place cartridge in analyzer
- Push bead into reagents
- Insert sample into cartridge

Performance Parameters
- Accuracy:The accuracy is tested with the reference material as samples, and the relative deviation of the determination results of the analyzer should be within the range of ±8%.
- Precision:Samples with a concentration of 4.0%-6.5% are detected, and the sample coefficient of variation (CV) of the test results is not more than 3%.
- Linearity:In the range of 4%-16% of HbA1c, the linear correlation coefficient r of the test results should be no less than 0.9900.
- Stability:Within 8 hours after the instrument is turned on and stable, the same normal sample is tested, and the relative deviation of the test results is not more than ±3%.
- Permission: The authority is the first-level authority, and the login authority is the first-level authority