HbA1c Test Analyzer LD-100 Hemoglobin Testing Analyzer POCT One-Step HbA1c Testing Analyzer
Intended Use
LD-100 POCT HbA1c Analyzer is intended for the percent determination of hemoglobin A1c (HbA1c) in human whole blood. Which is a semi-automatic instrument used to analyze glycated hemoglobin and it's suitable for the detection of HbA1c in medical units at all levels.
Specifications
Model | LD-100 |
Method | Fluorescent Immunoturbidimetry |
Operating Type | One-Step, fully automatic |
Measurement Time | < 3 minutes |
Sample Type | Whole blood |
Sample Volume | 4ul |
Data storage | 7000 |
Weight | 800gs |
Measurement Range | HbA1c 4.0~15%(NGSP Unit) eAG 68-385mg/dl(mmol/l) |
Operating Condition | Temperature:15~35℃ Humidity:15~80%R.H |
Storage Condition | Analyzer:10-60℃; Test kit: 2~8℃ |
Power connection | Detachable power supply cord |

PRINCIPLE
- Uses the (borate affinity) fluorescent immunoturbidimetry
- Push a mixture containing a fluorescent borate conjugate into an integrated test cup and mix to dissolve sufficiently.
- The blood samples in the sample collector are then pushed into the buffer and mixed to release the hemoglobin from the red blood cells.
- The total concentration of hemoglobin depends on the initial reduced fluorescence signal.
- Fluorescent borate conjugates bind to glycosylated hemoglobin, and glycosylated hemoglobin is determined by detecting the decrease of fluorescent active ingredients.
- The percentage of glycosylated hemoglobin in total hemoglobin can be determined.
Feature
- Simple Operation Step, One-Step Testing Process
- Protable, Light and Small,Easy to carry
- Fast test speed,Quick Recation
- Battery Integrated
- NGSP And IFCC Certificated
- Cv<3%, Excellent Correlation With HPLC
- RFID For Batch Management
- Color Touch Screen, Easy To Read
Reliable
IFCC calibrator | Normal | Abnormal | Normal | Abnormal |
Mean HbA1c% | 5.4 | 9.6 | 5.3 | 9.2 |
CV (%) | 2.5% | 1.9% | 2.4% | 1.8% |
Use Step
- Place cartridge in analyzer
- Push bead into reagents
- Insert sample into cartridge

Performance Parameters
- Accuracy: tested with the reference material as samples, and the relative deviation of the determination results of the analyzer should be within the range of ±8%.
- Precision: Samples with a concentration of 4.0%-6.5% are detected, and the sample coefficient of variation (CV) of the test results is not more than 3%.
- Appearance:All fasteners should not be loosened. Various adjusting elements can rotate flexibly and function normally. Readouts should be clear on the display screen.
- Linearity:In the range of 4%-16% of HbA1c, the linear correlation coefficient r of the test results should be no less than 0.9900.
- Stability:Within 8 hours after the instrument is turned on and stable, the same normal sample is tested, and the relative deviation of the test results is not more than ±3%.
- Permission: The authority is the first-level authority, and the login authority is the first-level authority.
Certificated