Hematology Reagent For Orphee Mythic 18 Analyzers High Performance Hematology Analyzer Reagents
Intended Use
To be used for blood cell counting, sizing, and WBC differentiation, hemoglobin determination on Orphee Mythic 18 hematology analyzers. This product is for In Vitro Diagnostic Use only.
PRECAUTIONS
- Re-calibration is recommended when changing the origin of the reagents.
- It is not recommended using different brands of solutions together on one unit of analyzer.
- Do not ingest. Avoid contact with skin and eyes.
- Skin contact: Flush with plenty of water.
- Eyes contact: Flush with plenty of water. Obtain immediate medical attention.
- This kit is for professionals and in vitro diagnostic use only.
Procedures
- Remove all the reagent aspiration tubes from the currently used reagent boxes or bottles.
- Wipe the tube’s surface with a lent free tissue, and then rinse with distilled water. Put the tubes into the LABNOVATION reagent boxes or bottles.
- Run at least 50 backgrounds. Check for the acceptable values. At first high background counts can be seen, that should lower to normal after a few runs.
- Run all three levels of a dedicated commercial control. Check for the acceptable values.
- Recalibrate the unit if necessary.
Specification
L-0602020 | MT-DILUENT | 20 L |
L-0602010 | 10 L |
L-0603010 | MT-LYSE | 1 L |
L-0603005 | 500 ML |
L-0604010 | CLEANER-MT | 1 L |
KIT COMPOSITION
MT-DILUENT | NaCl≤9.4%, Buffer≤1.6%, Anti-Microbial Agent≤1.0% |
MT-LYSE | Quaternary Ammonium Salt<10%, Sodium Lauryl Sulfate<1.5% |
CLEANER-MT | Buffer<0.3%,Proteolytic Enzyme<0.2% |
Advantage
- One source for all brands of analyzer
- Reagents for both 3-parts and 5-parts
- Ready-to-use final products and bulk OEM reagents
- ISQ13485:2003 and ISO9001:2008 certified
- CE marked, FDA cleared
- Well-established brand used by thousands of lab in 95 countries
- High quality, high performance and high fidelity of correlations with original brand.