Hematology Analyzer Reagents Hematology Reagent For CD3200 Analyzers High Quality Rreagents
Intended Use
To be used for blood cell counting, sizing, and WBC differentiation, hemoglobin determination on ABBOTT CELL-DYN Ruby hematology analyzers. This product is for In Vitro Diagnostic Use only.
Procedures
- Using the function keys press the following menus:
MAIN-SPECIAL PROTOCOLS-REAGENT RESERVOIRE-EMTY DILUENT
- Remove the reagent inlet tube from the Diluent.
- Wipe the tube’s surface with a lent free tissue, and then rinse with distilled water.
- Put the reagent tubes into the boxes and/or bottles of Labnovation reagents.
- Using the function keys press the following menu:
FILL DILUENT
This step will fill the reagent system with Labnovation reagent.
- Repeat steps 1-5 with Lyse and Sheath.
- Run at least 5 backgrounds. Check for the acceptable values.
- Run all three levels of a dedicated commercial control. Check for the acceptable values.
- Recalibrate the unit if necessary. (MAIN-CALIBRATION-ENTER-FACTOR)

Precaution
- Re-calibration is recommended when changing the origin of the reagents.
- It is not recommended using different brands of solutions together on one unit of analyzer.
- Do not ingest. Avoid contact with skin and eyes.
- Skin contact: Flush with plenty of water.
- Eyes contact: Flush with plenty of water. Obtain immediate medical attention.
- This kit is for professionals and in vitro diagnostic use only.
- Never put distilled water on the analyser (except clean for shipping menu), otherwise the pneumatic unit can be damaged seriously.
Specification
L-03H79-01 | Diluent/Sheath | 20 L |
L-03H78-38 | WBC Lyse | 3.8 L |
L-03H80-01 | HGB/NOC Lyse | 3.8 L |
33-40-E001 | 33-40-E001 | 50 ML×2 |
KIT COMPOSITION
Diluent/Sheath | NaCl≤9.4%, Buffer≤1.6%, Anti-Microbial Agent≤1.0% |
WBC Lyse | Surfactant, Lauroyl Sulfate, Ammonium salt and Buffer |
HGB/NOC Lyse | Quaternary Ammonium Salt<10%, Hydroxylamine Salt<3.5% |
EZ-Cleaner | Subtilisin<5.0% |
APPLICABLE INSTUMENT
- For use with CD3200 hematology analyzers.
Advantage
- One source for all brands of analyzer
- Reagents for both 3-parts and 5-parts
- Ready-to-use final products and bulk OEM reagents
- ISQ13485:2003 and ISO9001:2008 certified
- CE marked, FDA cleared
- Well-established brand used by thousands of lab in 95 countries
- High quality, high performance and high fidelity of correlations with original brand.
Storage
- Store the product at 5-35℃ in a dust free place with good ventilation. Temperature outside the range may damage the product.
- The open vial life of the product is 30/60 days. Unused reagents should be discarded after 30/60 days. Do not mix residual with a new carton.