Labnovation Technologies, Inc.

Labnovation Technologies,Inc.

Manufacturer from China
Active Member
4 Years
Home / Products / HPLC HbA1c Analyzer /

Accurate Test HPLC HbA1c Analyzer HbA1c Levels Detection For Diabetes

Contact Now
Labnovation Technologies, Inc.
City:shenzhen
Country/Region:china
Contact Person:MissAimee li
Contact Now

Accurate Test HPLC HbA1c Analyzer HbA1c Levels Detection For Diabetes

Ask Latest Price
Video Channel
Brand Name :LABNOVATION
Model Number :LD-560
Certification :IFCC, NGSP, CE
Place of Origin :Shenzhen, China
MOQ :1 Unit
Price :Negotiation
Supply Ability :100 Units/Day
Delivery Time :10-15 working days
Packaging Details :Carton Size: 480mm×480mm×615mm, Gross Weight: 28kgs
Name :HPLC HbA1c Analyzer Detection of HbA1c Levels for Diabetes
Model :LD-560
Principle :HPLC (High Performance Liquid Chromatography)
CV :<2%
Test Speed :90s/test
Sample types :Whole blood or Pre-diluent sample
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

HPLC HbA1c Analyzer Detection of HbA1c Levels for Diabetes


  • The full automatic analyzer provides quantitative measurement of the hemoglobin A1c(%HbA1c) levels using small blood sample(5ul).
  • This can help reduce the risk of diabetes complications
  • HbA1c levels is measured by HPLC method, designed for fast, easily, accurate test.


Specification

Model LD-560
Methodology High-Performance Liquid Chromatography (HPLC)
Test Modes Fast Mode, Variant Mode
Test Range 3% - 18%
Precision (CV) <2%
Parameter HbA1c, HbF, HbA1a, HbA1b, HbA0, HbA2(Optional)
First Sample Result 4 Mins
Test Speed Fast Mode: 90 Seconds/Test
Variant Mode: 140 Seconds/Test
Sample Type Whole blood, Pre-diluted blood
Sample Volume Whole blood: 5μL
Pre-diluted blood: 150μL
Sample loading capacity 20
Photometer 420nm
Column Capacity 3000T
Filter Capacity 600T
Display 10“ LCD touch screen
Reagent Kit Eluent A, Eluent B, Eluent C, Hemolysis, Chromatography Column
Storage 5000 Sample Results
Connection interface USB, LAN, LIS Compatible
Printer Thermal Printer

System operation

environment

Temperature 15-30℃
Humidity 30%-85%
Power AC 100-240V 50/60HZ 300VA
Dimensions 378mmx380mmx510mm
Weight 28Kgs


HPLC Methodology
The mobile phase in the solution reservoir enters the injector via the HPLC pump, where it is mixed with the sample.Each component of the mixture passes through the chromatographic column at different rates, and each component flows out of the chromatographic column sequentially due to the different adsorption capacities between each component and the stationary phase in the chromatographic column.Then, the component concentration is converted into an electrical signal and transmitted to the HPLC software of the computer, and the generated waste liquid enters the waste liquid bottle. After the operation is completed, the chromatogram can be obtained in the HPLC software.


Positioning

Accurate Test HPLC HbA1c Analyzer HbA1c Levels Detection For Diabetes


Performance

* HPLC- High Correlation with import HPLC

* CV<2% - Accurate and Stable

* Full Automatated - Automated sample loading and barcode scanning
* Quality & Performance - IFCC & NGSP Certificate


Contents

Main Parts

The analyzer consists of an auto sample injection module, a liquid chromatography separation module, a colorimetric detection and control module, a barcode scanning module, and a display printing module.


Reagents and Consumables

The analyzer is recommended to be used in conjunction with the Glycosylated Hemoglobin Test Reagent Kit (high performance liquid chromatography (HPLC method)) produced by our company, which includes the following components:
Eluent A/Eluent B/Eluent C(Variant Mode) Adjusts the pH of the solution for the elution of glycosylated hemoglobin and non glycated hemoglobin.
Hemolysis Reagent Used to dissolve red blood cells and release hemoglobin.
Chromatography Column Used for separation and elution of glycosylated hemoglobin and non glycated.
Hemoglobin Filter Used to initially filter the sample after hemolysis.
Paper Rolls Prints a report of the test results.
Accurate Test HPLC HbA1c Analyzer HbA1c Levels Detection For Diabetes


Certificate

Accurate Test HPLC HbA1c Analyzer HbA1c Levels Detection For Diabetes

Clover A1c : HbA1C Analyzer

  • The full automatic analyzer provides quantitative measurement of the hemoglobin A1c(%HbA1c) levels using small blood sample(4ul).
  • This can help reduction the risk of diabetes complications
  • HbA1c levels is measured small sized monitor, designed for fast, easily, accurate test.

Specifications

  • Principle : Boronate affinity Method : Spectrophotometry
  • Test Range : 4~14 %
  • Measuring time : 5 min (HbA1c) / 3,5,9 sec (Glucose)
  • Operating Temperature : room temperature(17-32 °C , 63-90 °F )
  • Sample : Capillary Whole Blood, Venous Blood with anticoagulant 4ul
  • NGSP certified & IFCC traceable
  • CE & FDA certified