Labnovation Technologies, Inc.

Labnovation Technologies,Inc.

Manufacturer from China
Active Member
4 Years
Home / Products / POCT HbA1c Analyzer /

800g POCT HbA1c Analyser Fully Automated For Glycosylated Hemoglobin Detection

Contact Now
Labnovation Technologies, Inc.
City:shenzhen
Country/Region:china
Contact Person:MissAimee li
Contact Now

800g POCT HbA1c Analyser Fully Automated For Glycosylated Hemoglobin Detection

Ask Latest Price
Video Channel
Brand Name :LABNOVATION
Model Number :LD-100
Certification :IFCC, NGSP, CE
Place of Origin :Shenzhen, China
MOQ :1
Price :Negotiation
Payment Terms :D/A, D/P, T/T
Supply Ability :100/Day
Delivery Time :15-20 working days
Packaging Details :Carton
Name :Glycosylated Hemoglobin Detection Analyzer POCT HbA1c Analyzer Fully Automated
Principle :Borate Affinity Method
Sample types :Whole blood
Sample Volume :4μL
Dimension :200mm(L)*130mm(W)*120mm(H)
Weight :800g
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

Glycosylated Hemoglobin Detection Analyzer POCT HbA1c Analyzer Fully Automated Analyzer

The Glycosylated Hemoglobin Detection Analyzer POCT HbA1c Analyzer Fully Automated Analyzer is a small, portable and fully automated point-of-care system forHbA1c test. The system delivers rapid tests of HbA1c.
The Glycosylated Hemoglobin Detection Analyzer POCT HbA1c Analyzer Fully Automated Analyzer may be used as a basic test system where the user purchases the Analyzer and manually records the results, or as an advanced system where a suite of accessories enable automatic printing of results, input of patient and operator identification numbers, and full connectivity to a PC. In this manner, the system
is highly versatile and can be tailored to specific user needs.


Feature

Boronate affinity Method: The HbA1c assay has no interference from common Hb variants

Ease of use: One-step operation the instrument is ready for use.

• Less than 3-minute test time

• A compact, lightweight instrument, Compact Size - To Minimize Space Requirements.

• With no need for bulky liquid reagents, LD-100 also free up more space in laboratories, clinics, blood banks, and other situations.

• Traceable to the IFCC Reference Method for HbA1c;Results reporting unit: IFCC units (mmol/mol) and NGSP units (%%).


Reliable

IFCC calibrator Normal Abnormal Normal Abnormal
Mean HbA1c% 5.4 9.6 5.3 9.2
CV (%) 2.5% 1.9% 2.4% 1.8%


Performance Parameters

  • Accuracy: Is tested with the reference material as samples, and the relative deviation of the determination results of the analyzer should be within the range of ±8%.
  • Precision:Samples with a concentration of 4.0%-6.5% are detected, sample coefficient of variation (CV) of the test results is not more than 3%.
  • Linearity:In the range of 4%-16% of HbA1c, the linear correlation coefficientrof the test results should be no less than 0.9900.
  • Stability:Within 8 hours after the instrument is turned on and stable, the relative deviation of the test results is not more than ±3%.
  • Functions of the analyzer:The function of report output in English; Abnormal alarm function.


Certificate

800g POCT HbA1c Analyser Fully Automated For Glycosylated Hemoglobin Detection800g POCT HbA1c Analyser Fully Automated For Glycosylated Hemoglobin Detection

Inquiry Cart 0